| SETUP
IN CCTC
PREPARATION
- Before
removing the set from the package, check all luer lock connections
to ensure they are tight. In particular, check the access and
return connections, the connections to all spikes and the connections
at the dearation filter. Connections are loosely connected to
facilitate gas sterilization.
- Prepare
solutions. Remember to add KCl to dialysate per protocol. If dialysate
is not being administered, KCl should be ordered to replacement
solutions per protocol.
- Standard
CCTC Setup:
- Predilution
replacement administered via Pre Blood Pump (PBP)
- Replacement
pump programmed to deliver POST DILUTION replacement (usually
at 200 ml/hr) to prevent clotting in the dearation chamber
- Citrate
Administration Only
- Administer
citrate via PBP
- Administer
all replacement fluid (e.g., 1000 ml/hr) POST DILUTION
via replacement pump.
- Always
turn the machine "OFF", then back "ON" before
setting up a new filter (even if you restarting the same patient).
This will ensure you have cleared any alarms from the previous
run.
- Do not turn
the heater on until after the circuit is connected to the patient
and the heater tubing is filled with blood. When bicarbonate containing
solutions are heated, CO2 gas is produced, increasing the number
of bubbles.
SET
PREPARATION
- Always
setup in "CVVHDF" mode.
- Remember
to connect the heater tubing.
- Position
the green marking on the heater tubing at the front of the
heater and wrap from front to back.
- Pull
and stretch the tubing during positioning to reduce the chance
of kinking when the cover is replaced.
- Have
a second person assist you as you open up the blue connection
in the return tubing (below and to the left of the dearation
chamber) and connect the heater tubing. There is only one
way the heater tubing can be connected (male to female). It
is normal for the return tubing to cross behind the filter
and again underneath the heater.
PRIMING
- Be
sure that the connections are tight, especially at the return
protector (the filter above the dearation chamber), the access
and return connections, and the luer lock connection on the spike
inserted into the prime collection bag.
- If
heparin is not being administered, prepare a 20 ml syringe with
a LUER LOCK connection with 0.9 NaCl and install into the syringe
pump.
- Hang a bag
of normal saline on any pump that is not being used (e.g., if
dialysate had not been ordered, hang saline on the dialysate pump
during priming and set the dialysate flow rate at "0".
- In the last
minute of priming, you can gently flick or wiggle the loop in
the heater tubing to faciliate removal of air bubbles. Small air
bubbles do not need to be removed.
- It is normal
to have air in the effluent pod or a few inches of air in the
effluent tubing. You DO NOT need to prime a second time; this
air will be eliminated during the Prime Test.
- Air bubbles
in the blood side will be collected in the dearation chamber.
- At the end
of priming, check the dearation chamber and adjust the level if
required.
- Following
priming, an air pocket will sometimes collect in the dearation
chamber below the mesh filter, displacing fluid into the tubing
above the dearation chamber.
- Ensure
that the
- Always
check the dearation chamber carefully; it can be difficult
to differentiate fluid from air.
- If fluid
has risen into the tubing above the chamber, lower the fluid
level by choosing "adjust level", and using the
DOWN arrows to drop the fluid level below the dearation chamber.
After the level is dropped, select the UP arrows and move
the fluid level to the desired location.
- It can
be difficult to differentiate fluid from air. Dropping the
level , appearing as though fluid is in the chamber.
- DO
NOT do any additional priming without hanging a second bag
of priming solution or the bag will run dry during the prime
test (the prime test using the remaining 150 ml of the 1 Litre
bag of solution).
CONNECTING
TO PATIENT
At the end of
priming, select "continue" and enter flow rates prior
to connecting the circuit to the patient.
Initate
treatment with fluid removal set at "0" and blood flow
rate at 150 ml/min.
Prior to connecting
the circuit to the patient, switch the blue return line and the
effluent line. (The effluent line should be moved to the effluent
bag, and the return line should be moved to the "Y" connector
on the priming "Y". Clamp the access and return lines
and move the priming bag close to the patient's access site to facilitate
connection.
Disconnect the
access tubing from the "Y" connector and add a 3-way stopcock.
Prime a regular IV tubing set with normal saline and connect to
the stopcock.
Prepare the
limbs (check for clots with a 3 cc syringe and flush with saline).
Instill the
heparin bolus (if ordered) into the access limb and connect the
stopcock at the end of the access line to the access limb.
Conect the return
line to the return limb (blue to blue).
Do
not manually prime without hanging a second bag of priming solution
(use the same priming solution used for the first prime, e.g., add
heparin if ordered). You will run out of saline during the Prime
Test if you try to manually prime with the intial bag.
YOU
MUST CONNECT THE PATIENT WITHIN 10 minute of priming. If more than
10 minutes elapses, repriming must be performed again.
ACCESS
Historically,
early circuits removed blood from arterial access sites and returned
the purified blood via a venous catheter. This promoted blood flow
through the filter by utilizing the patients own arterial to venous
blood pressure gradient.
Although
arterial-to-venous access sites are still used in patients with
end-stage renal failure, the blood flow pumps on modern hemodialysis
machines eliminate the need for arterial-to-venous pressure gradients.
Consequently, venous-to-venous circuits are often used. An example
of an arterial-to-venous access site is a fistula. Fistulas are
created surgically using graft material to connect a limb artery
limb directly to a vein. They take several months to "mature"
or dilate sufficiently to enable use of the fistula for dialysis
access. Because they are under the skin, fistulas reduce the risk
for bleeding and infection when long term therapy is required. Filter
patency can be confirmed by listening with a stethoscope for a "bruit"
(created by the rapid flow of blood though this large circuit) or
by feeling a tingling sensation known as a "thrill". Never
place a BP cuff or tourniquet above a fistula as this can reduce
the flow through the circuit and lead to clotting.
 |
 |
| Fistula |
Double
Lumen Venous Catheter |
In
critical care, temporary double lumen venous dialysis catheters
are the most common form of access. They can be inserted quickly
at the bedside and used immediately. "Perm" catheters
are double lumen venous catheters that are designed for longer indwelling
use. They are used more frequently in patients with chronic renal
failure and may be used as a bridge until a surgically created fistula
is ready for use.
Dialysis
catheters are easy to differentiate from regular intravenous lines
by their red and blue hubs. The red lumen denotes the side of the
venous catheter that is used to pull blood from the patient, and
is referred to as the access lumen. The blue lumen is the return
site and is used to reinfuse the patient's blood after it passes
through the dialysis filter. If an adequate flow rate cannot be
achieved by removing blood from the access side of a catheter, the
catheter limbs can be reversed. Reversal of the limbs does produce
a small reduction in clearance due to recirculation.
Cautionary
Note:
A double lumen venous dialysis catheter can technically
be used as a central venous infusion site, however, to ensure the
line remains patent for subsequent dialysis treatments, and to reduce
the risk for infection, it is preferred that these catheters are
used only for dialysis. If in a life-threatening situation it is
the only vascular access available, it can be used as a central
line, HOWEVER, it should always be assumed that the catheter contains
heparin. The amount of heparin used to block the catheter limbs
varies between dialysis programs. Typical concentrations are 5,000
- 10,000 units per mL. Because each lumen contains a volume of between
1.2 and 2 ml, limbs could contain up to 40,000 units of heparin!
For this reason, ALWAYS assume that heparin is in the limb of any
dialysis catheter. Always aspirate at least 5 mL of blood from EACH
lumen prior to using the catheter.
Beginning
April 1, 2005, we will begin using citrate solution to block all
dialysis lines in CCTC (instead of heparin). Despite this change
in practice, ALWAYS aspirate blood from both limbs before using
a dialysis line for intravenous therapy.
PRINCIPLES
Hemodialysis
employs the principles of diffusion, hemofiltration and convection,
using an external filter to create an artificial nephron unit.
Recall
the normal nephron unit:
|
A:
Renal arteriole
B: Afferent arteriole
C: Efferent arteriole
D: Bowman's Capsule
E: Glomerulus
F: Proximal Tubule
G: Loop of Henle
H: Distal Tubule
I: Collecting Tubule
|

|
Diagrams:
http://coe.fgcu.edu/faculty/greenep/kidney/nephron.htm |
The
diagram above depicts one nephron unit. Each kidney has approximately
one million of these microscopic units. They collectively maintain
water, electrolyte, waste and acid-base balance.
Arterial
blood flows from the renal artery, branching into smaller divisions
known as arterioles. Branches of the arterioles eventually carry
blood into small "containers" called Bowman's Capsules,
located in the cortex of the kidney. The arteriole that Arrives
at the Bowman's Capsule is called the Afferent
arteriole. The blood then flows into a specialized capillary (located
inside the Bowman's Capsule), called the GLOMERULUS. Any blood remaining
at the end of the glomerulus Exits the Bowman's
Capsule via the Efferent arteriole.
The
afferent arteriole is larger in diameter than the narrow efferent
arteriole. This arrangement provides a high rate of blood flow into
the glomerulus, but a high resistance to blood leaving the glomerulus.
This structural difference from other capillaries ensures that the
pressure within the glomerulus will be twice that of other capillaries.
The increased glomerular pressure creates a hydrostatic force to
promote filtration. The wall of the glomerulus is semi-permeable,
allowing small and mid sized molecules to freely pass across the
glomerular membrane and into the Bowman's Capsule. The increased
hydrostatic pressure increases the movement of water across the
membrane, dragging solutes or particles along in the process. The
"washing" of solutes across the membrane with this large
flux of water is known as convection. Because proteins are too large
to fit across normal glomerular membranes, the protein remains in
the blood and returns via the efferent arteriole. This prevents
protein from appearing in the urine.
In
order to eliminate all of the wastes that we want to remove, we
have to filter very large volumes of water across the glomerulus.
About 1200 ml per minute of filtrate is produced. By the time enough
water has been moved across the membrane to wash out all of the
surplus waste products, over-removal of water, glucose, electrolytes
and other substances has occurred. Consequently, water and solutes
must be reabsorbed from the filtrate in the tubules to maintain
normal intravascular concentrations. Solutes and water are reabsorbed
into capillaries that are wrapped around the tubules, called peritubular
capillaries. The efferent arterioles become the peritubular capillaries,
and are also responsible for perfusing the kidney. Surplus solutes
can also be secreted into the tubules for elimination.
Reabsorption
of needed water and solutes and secretion of surplus solutes continues
along the renal tubules. The collecting tubule at the end of the
tubule system contains the final filtrate. Any water or solute remaining
in the collecting tubule is eliminated in the urine.
Artificial
Kidneys
The dialysis filter is referred to as an artificial kidney. Blood
is pulled from the patient and carried into the filter. Once inside,
the blood travels through many tiny tubules called hollow fibers.
Water and solutes can pass across the semi-permeable membrane between
the blood and the fluid that surrounds the hollow fibers. Any fluid
or solutes that enters the filter canister will be drained out as
waste.
 |
| Schematic
of dialysis filter (artificial kidney) |
Note
how the dialysis filter has structural similarities to the nephron
unit. Blood arrives at the filter via the access tubing (afferent
arteriole). Blood enters the small hollow fibers within the filter
(glomerulus). Water and solutes diffuse across the semi-permeable
membrane of the hollow fibers and collect in the canister (Bowman's
Capsule). Collected fluid (filtrate or effluent) is then removed
via the drainage tubing (collecting tubule). Blood that remains
in the hollow fibers is returned to the patient via the return side
of the filter (efferent arterial).
Although
similarities exist between the nephron unit and the artificial kidney,
the artificial kidney has limited capabilities. In the nephron unit,
filtered water and waste enters the proximal tubule. Because the
nephron unit removes significantly more water and solutes than needed,
most of the water and electrolytes that enter the tubule system
are reabsorbed.
Unlike
the nephron unit, the artificial kidney cannot reabsorb water or
solutes that enter the filter canister Any filtrate that enters
the filter canister will be removed via the drainage tubule. Consequently,
one of the differences in the artificial kidney is the absence of
the proximal tubule, loop of henle and distal tubule where water
and solute reabsorption and secretion occurs. Thus, the drainage
tubule that exits the filter is similar to the collecting tubule
of the nephron unit, not the proximal tubule. To compensate for
the inability to reabsorb water and solutes following removal from
the blood, the artificial kidney is manipulated to restrict the
actual removal to only surplus water and wastes. This is done by
adjusting dialysis solutions and ultrafiltration rates. If more
water or solutes are removed than desired, they may need to be given
back via intravenous infusions.
The
artificial kidney does not replace other important kidney functions,
including stimulation of red blood cell production (erythropoietin),
blood pressure and sodium regulation (renin) and calcium uptake
by the GI tract (vitamin D synthesis). The nephron normally traps
and recycles bicarbonate to maintain acid base balance. Bicarb is
given to patients during hemodialysis to compensate for bicarb deficits.
The
principles used during hemodialysis are reviewed below:
DIFFUSION
Diffusion
is the movement of particles (solutes) across a semi-permeable membrane.
Diffusion is the movement from the side with the highest concentration
of particles, to the side with the lowest concentration.
DIALYSIS
FLUID (DIALYSATE):
Dialysate is the fluid that is pumped into the filter canister,
surrounding the hollow fibers. The concentration of solutes in the
dialysis fluid determines diffusion gradients. The removal of surplus
solutes from the blood is achieved by infusing dialysate fluid that
contains a lower solute concentration than the serum concentration
(e.g. dialysate does not contain urea or creatinine).
To
maintain normal serum electrolyte levels, dialysate fluid contains
sodium, chloride and magnesium levels that are equal to serum concentrations
(thus, removal of these electrolytes should only occur if the blood
level exceeds normal serum concentrations). In renal failure, potassium
is often high at the start of a treatment, therefore, we may begin
dialysis with a low concentration of potassium in the dialysate.
Because potassium is easily removed during dialysis, and continued
dialysis will be required to ensure removal of other wastes such
as urea and creatinine, potassium concentrations in the dialysate
often require upward adjustment as the potassium level in the blood
falls. Although in theory, potassium levels should not fall below
4 mmol/L in the serum if the dialysate contains 4 mmol/L, a number
of factors influence serum potassium levels in critical care. Insulin
therapy and the use of sympathomimetic drugs promotes the movement
of potassium from the blood into the cells. This can lower serum
levels. Additionally, potassium loss through the GI tract can increase
the potential for hypokalemia. Low magnesium levels will also suppress
the serum potassium levels, therefore, magnesium deficits should
be replaced as needed. Additionally, high hemofiltration rates can
lead to additional potassium clearance. Potassium levels must be
monitored closely and adjusted to maintain normal serum concentrations.
In
renal failure, serum bicarbonate levels are generally low, therefore,
a source of bicarbonate is added to the dialysate to facilitate
diffusion of bicarbonate into the blood. Lactate based formulas
provide one source (e.g., Gambro's LG formulas). Higher concentrations
of lactate in the dialysate promote diffusion into the blood. If
hepatic function is normal, lactate is quickly converted to bicarbonate
in the liver. Prisma(TM) and
Prismaflex(TM) both use premixed
bags of sterile dialysate. Lactate based preparations have a long
stability, making them less expensive to prepare. Because bicarbonate
is only stable for a short period in solution, it must be added
to the dialysis bags before using.
If the patient is unable to convert lactate to bicarb at a rate
that is fast enough, serum lactate levels will rise. This occurs
in both hepatic insufficiency and shock states were the patient
already has excess lactate production due to anaerobic metabolism.
In these instances, a bicarb containing product is used to deliver
bicarbonate (e.g. B.O or Normocarb).
If
1 L of dialysate is administered per hour, one L of dialysate fluid
will collect in the drainage collection bag per hour. This will
be in addition to any fluid removed; dialysate doesn't normally
cross into the bloodstream.
Concentration
gradients play a major role in diffusion. These will be explored
further in the discussion on clearance. The other factor that influences
diffusion is the type of filter used. Diffusion of solutes cannot
occur across a concentration gradient if the pore size is too small
to permit passage.
ULTRAFILTRATION
Ultrafiltration is the movement of water across a semi-permeable
membrane because of a pressure gradient (hydrostatic, osmotic or
oncotic). The increased blood pressure in the glomerulus creates
a favourable driving pressure to force water across the glomerular
membrane.
Blood
pressure within the hollow fibers is positive, while the pressure
outside the hollow fibers is lower. Increased negativity can be
generated outside the hollow fibers by the effluent pump by either
increasing the fluid removal rate, or by increasing the replacement
flow rate. The difference between the blood pressure in the hollow
fibers and the surrounding pressure is the TransMembrane Pressure
(TMP). The TMP determines the ultrafiltrate production.
Different
filter membrane properties can produce different ultrafiltration
rates at a constant TMP. A filter that is more permeable to water
will allow more water to travel across the membrane at a given TMP.
A filter with a high permeability to water is called a high flux
membrane.
HEMOFILTRATION
In
hemodialysis circuits, pulling large volumes of water across the
semi-permeable membrane creates a convective current that "drags"
additional solutes. While diffusion is effective at removing most
small molecules, convection enhances the removal of small and mid-sized
molecules. Thus, convection can be added to hemodialysis therapy
to enhance solute removal. To prevent hypovolemia, any water removed
during hemofiltration must be returned to the blood before it reaches
the patient. This is called "replacement" fluid. Hemofiltration
rates of 1 L/hr mean that one liter of fluid is removed from the
patient's blood and eliminated in the drainage fluid AND 1 L of
replacement fluid is returned to the circuit before it reaches the
patient. We set hemofiltration rates by adjusting replacement rates.
Any fluid removed during hemofiltration is given back to maintain
a net neutral fluid balance. Replacement fluid must be sterile intravenous
fluids with concentrations of electrolytes similar to plasma.
For
example, if the CRRT therapy includes a hemofiltration rate of 1
L per hour, and the fluid removal is set at 200 ml per hour, 1200
ml will be pulled from the patient and introduced into the drainage
collection bag each hour. Because the 1 L of hemofiltration is replaced,
the net fluid removed is 200 ml. Whether hemofiltration is used
or not, the net fluid removed is equal to the fluid removal setting.
PREDILUTION
VERSUS POSTDILUTION HEMOFILTRATION
Replacement
fluids can be returned either pre or post filter. This is referred
to as predilution or post dilution sets. Predilution means that
the replacement solution is returned to the blood before it reaches
the filter, diluting the blood in the hollow fibers. Postdilution
means that the replacement fluid is returned to the blood after
the filter (but before the return side of the access catheter).
Predilution dilutes the blood in the filter, reducing clotting.
Postdilution concentrates the blood in the filter, enhancing clearance.
CLEARANCE
Creatinine
is a byproduct of muscle protein metabolism that is completely filtered
by the glomerulus and 100% eliminated. None of the filtered creatinine
is reabsorbed from the tubules nor is any additional creatinine
secreted into the tubule lumen post glomerulus. This makes it the
best indicator of renal failure. Because it is completely eliminated
during normal renal function, measurement of creatinine clearance
is the best measure of glomerular filtration.
Urea is another byproduct of protein metabolism, however, it is
a byproduct of all protein metabolism (not just muscle protein metabolism).
It is filtered into the glomerular filtrate. Unlike creatinine,
a percentage of filtered urea is reabsorbed from the tubules. Consequently,
urea levels can become increased in the presence of a normal creatinine
level. For example, urea can increase due to increased urea production
(e.g., anabolic or catabolic states) or increased tubule reabsorption
of urea (e.g., due to dehydration). Creatinine only increases when
renal filtration decreases, or the production of creatinine becomes
so high that it exceeds glomerular filtration capabilities. Excessive
creatinine production can occur when significant muscle death has
occurred, for example in rhabdomyolysis.
Clearance
is the rate at which solutes are cleared from the body. Clearance
is abbreviated by the letter K. The clearance
(or K) of a solute is the volume of blood from which the substance
is completely removed per unit time (Gambro training manual). It
is calculated as follows:
K
= excretion rate of solute / blood concentration of solute
To
translate this to dialysis: if a dialyzer has the ability to clear
170 ml/min of urea at a blood flow rate of 200 ml/min, it means
that for every 200 ml of blood that flows through the filter, 170
ml will be returned urea free. The remaining 30 ml will have the
same concentration of urea as the blood entering the filter. The
200 ml of blood being returned each minute to the systemic circuit
will have significantly less urea than without dialysis, but will
still have to mix in with the systemic volume. Thus, blood must
continually circulate through the filter before the total systemic
level will begin to fall.
The
following formula can be used to calculate the clearance of a solute
in ml/min at the dialysis membrane. To calculate the rate of clearance
of a solute, the following formula can be used, where Q(blood)in
is the flow of blood into the filter, Q(blood)out is the flow of
blood out of the filter, C(blood)in is the concentration of the
solute in the prefilter serum and C(blood) is the concentration
of the solute in the post filter blood. Q(blood)in and Q(blood)out
are the same and equal to the blood flow rate.
|
| This
can be simplified to: |
 |
Example
below:
At a blood flow rate of 150 ml/min, with a prefilter creatinine
of .980 and post filter concentration of .343, the creatinine
clearance by the filter is 97.5 ml/min. This assumes that
no hemofiltration is being used. |
 |
FILTERS
Dialysis
membranes need to be efficient at clearing wastes, but must also
be biocompatible with human blood. Compatibility means that exposure
of blood to the dialysis membrane produces minimal of adverse effects.
Filter permeability is influenced by pore size, the number of pores
and the thickness of the membrane. Generally, high flux membranes
which have more or larger pores allow more solutes and ultrafitrate
to move across the membrane. Thinner membranes offer less resistance
to solute movement by decreasing the distance the solute must travel
across the membrane and also favours increased filtration.
Solutes
are pass through the membrane according to solute size. Imagine
taking a flour sieve and filling it with a mixture of sand, small
rocks and debris. Shaking up the contents would cause the smallest
particles to move towards the bottom, passing through the openings
easily. Particles would be filtered through according to increasing
size until you are left with particles to large to fit through the
sieve. Dialysis membranes act the same way, allowing small and mid
sized molecules to pass across the membrane, without the loss of
larger proteins. High flux membranes that have a larger pore size
increase clearance by allowing larger molecules to pass through
the membrane, and by allowing more ultrafiltrate flow. The standard
AN69 filter used with CRRT is a high flux membrane. Sieving properties
of a membrane describe the membrane's permeability to solutes during
ultrafiltration. Permeability of solutes decrease as the the molecular
size increases. The cut-off point for a membrane is defined by the
molecular weight where only 10% of the solute is filtered.
The
surface area of the membrane determines the available area for diffusion
and ultrafiltration. The internal volume of the dialysis filter
should be small enough to limit the amount of blood that is outside
of the vascular compartment at any given time. This volume is important
if the filter clots before blood can be returned to the patient.
Finally,
adsorption is the ability of larger solutes to adhere to the surface
of the dialysis membrane. AN69 filters used in CRRT have strong
adsorptive properties. Adsorption of mid sized molecules including
inflammatory mediators have been demonstrated by a drop in serum
concentrations following initiation of a new filter. The greatest
benefit appears to occur in the first few hours; once the filter
becomes saturated with proteins, further removal from the serum
is limited. While these proteins are too large to pass through the
filter and be removed in the filtrate, the removes the cytokines
from the blood by allowing the to collect (like a sponge) in the
filter.
TMP
is the pressure exerted on the dialysis membrane during operation
and reflects the difference between blood and fluid compartments.
A TMP above +350 mmHg will produe an advisory alarm. A TMP >
450 will produce a "TMP excessive" alarm. The amount of
increase and the rate of TMP increase contribute to the "Filter
is Clotting" alarm.
Filter
Pressure drop is another indicator of clotting. It is an indication
of the pressures in the hollow fibers of the filter. It will slowly
rise with filter use as the hollow fibers become filled with microscopic
clot. The amount and rate of increase determines the activation
of the "filter is clotting alarm".
DIFFUSION
Small
molecular weight solutes are easily removed by diffusion (dialysis).
The higher the concentration gradient, the higher the diffusion
rate. Solutes will move across a semipermeable membrane until the
two solute concentrations become equal.
As
solutes move into the dialysate fluid, the dialysate concentration
of the solutes increase, reducing the diffusion gradient. Once the
dialysate concentration of a solute becomes equal to the blood concentration,
diffusion stops. To maintain a high diffusion gradient, the difference
between the blood and dialysate concentrations must be maintained.
Clearance can be increased by higher dialysate or blood flow rates.
Increasing the dialysate rate maintains a low concentration of solutes
on the dialysate side by increasing their removal from the dialysate
fluid. Increasing the blood flow rate brings more solutes to the
filter, promoting continuous diffusion. The smaller the molecule,
the greater the clearance by dialysate/blood flow increases.
Although
higher blood flow rates will increase the rate of clearance, CRRT
circuits have limitations. The smaller filter size (compared to
hemodialysis circuits) limits the blood flow rates. Blood flows
can be increased substantially with hemodialysis, however, blood
flow rate adjustments are limited with CRRT.
While increased dialysate flow rates enhance the clearance of small
molecules, middle sized molecule clearance is more dependent upon
the size of the filter pores. The only way to increase the clearance
of middle sized molecules is to add convection (hemofiltration).
Optimal
solute clearance is produced when dialysate flow rates are approximately
double that of the blood flow rates. CRRT blood flow rates are typically
150 ml/min. A dialysate flow rate of 1 L per hour, provides a dialysate
flow of 16 ml/min. Increasing the dialysate flow will have a greater
effect than any increase in blood flow rates with CRRT.
Dialysate
flows countercurrent, or in the opposite direction to blood flow.
This promotes continual clearance by ensuring an adequate diffusion
gradient is maintained. Dialysate fluid is introduced at the return
end of the filter, where the serum concentration of solutes has
begun to fall (due to removal from the blood within the filter).
The dialysate fluid flows towards the access end of the filter where
the fluid drainage tubing is located. Diffusion of solutes along
the filter makes the concentration of wastes highest in the dialysate
at the access end of the filter. At the access end, the blood concentration
of the solute is highest , counterbalancing the rising dialysate
concentration.
HEMOFILTRATION
Dialysis
effectively removes small (e.g. electrolytes) and small to mid size
molecular weight solutes (e.g. glucose, urea, creatinine). The pore
size limits the ability to diffuse middle sized molecules. One way
to increase the clearance of all small and more of the mid sized
molecules is to pull large quantities of water across the semi-permeable
membrane, “dragging” additional solutes by convection.
Higher hemofiltration rates are of interest in critical care. Higher
pre dilution rates may be a successful alternative to anticoagulant
therapy, although, research is needed to examine this option. There
is also interest in the potential clearance of mid sized molecular
weight solutes including inflammatory mediators. In a European trial,
hemofiltration rates of 35 ml/kg/hr were associated with the best
survival rates. Although higher hemofiltration rates have been used
in CCTC, our current practice uses predilution therapies. In this
trial by Ronco, post dilution was used. The significance of high
hemofiltration rates using predilution replacement is not known.
While
increased ultrafiltration rates during hemofiltration help to remove
molecules too large to travel by diffusion, hemofiltration can also
lead to excessive removal of small molecules. Consequently, electrolyte
removal can be increased beyond that produced by the diffusion gradient
alone (e.g., despite a dialysate concentration of sodium that is
equal to normal serum levels, sodium levels can fall with high hemofiltration
rates).
Alternatively,
high infusion rates of replacement fluids containing 0.9 NaCl can
lead to hypernatremia. It can also increase chloride levels leading
to hyperchloremic acidosis (chloride and bicarbonate are both negatively
charged, increased chloride levels can cause a decrease in bicarbonate
to maintain anionic balance).
When
hemofiltration rates are high, careful monitoring is required to
maintain normal electrolyte balance. Replacement fluids may need
to be adjusted to keep serum levels within range. Alternatively,
intermittent boluses of electrolytes may be required.
THERAPIES
Original
continuous hemodialysis circuits required arterial to venous access
sites, because they did not utilize a blood pump to pull blood through
the filter. Consequently, they were referred to as CAV (Continuous
arterial-venous) circuits. Today's technology uses a blood flow
pump, therefore, most continuous circuits are CVV (continuous venous-venous).
SCUF
(Slow Continuous Ultrafiltration):
SCUF is the removal of water from the patient's blood as it travels
through the filter. Water removal is referred to as ultrafiltration.
SCUF is a therapy designed to only remove surplus water. The amount
of water removed is not sufficient to remove wastes.
CVVH
(Continuous Venous-Venous Hemofiltration)
CVVH is the removal of large amounts of water across the filter
membrane for the purpose of clearing wastes. When large volumes
of water are washed across the membrane, solutes are dragged along
with the water (convection). Hemofiltration is the removal of water
over and above the surplus water removed during ultrafiltration.
To prevent hypovolemia, water removed during hemofiltration must
be given back before the blood is returned to the patient. This
is referred to as replacement. CVVH is the use of replacement fluid
without dialysis fluid, plus or minus fluid removal.
CVVHD
(Continuous Venous-Venous Hemodialysis):
CVVHD is the infusion of dialysis fluid into the filter canister
The dialysis fluid (dialysate) surrounds the blood filled filter
segments. Solutes that are small enough to fit through the membrane
of the dialysis filter will move from an area of high concentration
to low concentration (diffusion). The dialysate determines the solutes
that will be removed. If we want to remove solutes, the concentration
in the dialysate is lower than the blood concentration. If we want
to give something to the patient, the concentration in the dialysate
is higher than the blood. CVVHD is the removal of wastes by diffusion
only, without the use of hemofiltration (replacement fluid). It
can be administered with or without fluid removal from the patient.
CVVHDF
(Continuous Venous-Venous HemoDiaFiltration):
CVVHDF is the use of dialysis AND hemofiltration.
Therapy will include the use of both dialysate and replacement fluids
and can be administered with or without fluid removal from the patient.
Return
to top
References:
Gambro Training Manual 1 and 2
Slides from Gambro Training package, reproduced with permission
Last
Update:
March 30, 2010.
|