|
PROCEDURE
SET-UP OF HEMODYNAMIC
MONITORING CIRCUIT IN CCTC |
- Assemble transducer pole clamp
- Open transducer kit
- Tighten Connections
- Prepare flush solution
- Prime the drip chamber
- Prime the air-fluid interface stopcock
- Prime the tubing
- Position stopcock for line insertion
|
- Replace vented caps
- Pressurize fluid
- Connect tubing to patient
- Level system
- Zero transducer
- Replace cap
- Examine waveform
|
|
PROCEDURE |
RATIONALE
FOR PROCEDURE |
|
1. |
Assemble and tighten
transducer clamp to IV pole. |
1. |
- The
transducer holder frees the user's hands to facilitate ease
in priming.
- The
pole clamp permits easy repositioning of the transducer to ensure
correct leveling during patient position changes.
|
|
2. |
Open
transducer kit and place the transducer firmly in any one
of the three brackets.
- Insert
until you hear a click.
- Position
the transducer so that the stopcock is above the transducer
and the flush device is below.
- The
tubing which leads to the patient catheter can be identified
by its hard non-compressible appearance. The tubing which
delivers the flush infusion is soft and compliant (like regular
IV tubing).
|
2. |
- Air
rises to the top; placement of the stopcock above the transducer
promotes easy clearance of air from the circuit.
- The
tubing which carries the pressure wave from the patient to the
transducer is hard and non-compliant to ensure that external
pressure (e.g. kinkage of the tubing) cannot influence internal
tube pressure and lead to false pressure recordings. The tubing
which delivers the flush is positioned pre transducer and therefore
has no influence on pressure measurement.
|
|
3. |
Tighten
all luer connections. |
3. |
Connections
are loose during packaging to facilitate gas sterilization process.
If left loose, pressure loss may lead to inaccurate readings or
backup of blood. |
|
4. |
Perform hand hygiene, don non-sterile gloves.
Open
premixed bag of heparinized saline solution.
- Invert
bag, placing medication administration port straight up towards
ceiling. Using 18 gauge blunt needle, puncture medication administration
port and eliminate all air from bag.
- Alternatively,
if premixed solution is unavailable, add 1,000 units of heparin
to one 500 mL bag of .9 NaCl.
- Document
heparin on medication administration record.
DO
NOT ADD HEPARIN IF PATIENT IS SUSPECTED OF HAVING HIT SYNDROME
OR HEPARIN ALLERGY. |
4. |
- In accordance with the MoHLTC 4 moments of hand hygiene and LHSC infection control policies in an effort to reduce risk of transmission of microorganisms and secretions.
- Air
will rise to the top. Keep both the port and needle at
exactly 90o to the floor to prevent air bubbles from
being trapped in the edges of the bag.
- Elimination
of air will prevent air bubbles from being pressurized into
the circuit.
- Air
entry to the circuit will interfere with pressure measurements
and can be difficult to clear from the line if it reaches the
transducer.
- Large
air bubbles entering the blood stream can cause air embolus.
|
|
5. |
Spike
the infusion bag with the IV tubing. Hang the infusion bag
on the hook inside the Tycos infusor. Do not pressurize the
system yet.
- Open
the roller clamp and squeeze the drip chamber slightly. Close
the clamp and release the drip chamber.
- The
goal when setting up the system is to have approximately
1/4 - 1/3 of the drip chamber filled with fluid.
- This
is enough to prevent air from entering the circuit, while
retaining visibility of the flow rate during flushing.
|
5 |
- Pressurization
prior to priming will promote small air bubbles to accumulate.
- As
soon as the system is pressurized, the level in the drip chamber
will rise higher; air in the top of the trip chamber allows
visualization of the rate of flushing.
|
| 6. |
Open
the roller clamp.
- Turn
the stopcock above the transducer "OFF" to the patient line
(the system will now be "OPEN TO AIR").
- Gently
pull the flush devise.
- Prime
the system until all air has been evacuated.
- Check
the transducer carefully to ensure that all air bubbles have
been evacuated.
- The
white cap over the stopcock is vented; you may flush through
it without removal.
|
6. |
- Gravity
versus pressurized priming minimizes the formation of bubbles.
- A
solid column of fluid that is free from air bubbles must cover
the transducer membrane. Bubbles will interfere with pressure
waveform transmission.
|
|
7. |
Turn
the stopcock above the transducer "OFF" to the vent port (air
elimination port).
- Check
that the patient stopcock at the end of the patient line is
closed to the blood drawing port.
- The
end cap is vented and can be flushed through without removal.
- Pull
the flush device to continue priming to the end of the patient
tubing.
|
7.
|
- Any
air bubbles in the patient tubing will dampen the transmission
of pressure waveforms.
|
|
8. |
Turn
the stopcock closest to the patient catheter off at a 45o
angle to the blood drawing port (half-way between two ports). |
8. |
- Turning
any 3 way stopcock at a 45o angle closes the stopcock
in all three directions. This will prevent the circuit from
dripping on the floor.
- When
pressurized, the flush device will allow 3 - 5 mL per hour into
the patient tubing unless the system is turned off.
|
|
9. |
Replace
vented caps on blood drawing and air elimination port with leur
lock dead-end cap. |
9. |
All
indwelling lines must have dead-end, leur lock caps to protect against
bleeding out or air entry into the system. |
|
10. |
Pressurize
the Tycos by turning the clamp on the back in the direction
shown for "CLOSE" (clockwise). Continue to tighten until
the arm swings freely. |
10. |
- The
bag must be pressurized to oppose the blood vessel pressure.
- Inadequate
pressure will cause blood backup, inaccurate pressure readings
and loss of line patency.
|
| 11. |
Connect
the end of the patient tubing to the patient catheter.
- Turn
the patient stopcock off to the blood drawing port.
|
11. |
- This
opens the circuit to patient monitoring.
|
| 12. |
Level system
- Position
the vent port (air elimination port) on the stopcock that
is located above the transducer in a position that is level
with the 4th intercostal space, midaxillary line (for any
arterial line, central line or pulmonary artery catheter).
|
12. |
- The
4th intercostal space, midaxillary line is the reference point
for monitoring cardiac pressures.
- When
both ends of the patient tubing are level (the transducer stopcock
and the heart are horizontal, the pressure that is exerted on
the transducer as a result of the weight of the fluid in the
tubing will be neutral.
- If
the transducer is lower than the patient's heart, the pressure
exerted by the fluid column will "run down towards the transducer",
falsely elevating the pressure reading.
- If
the transducer is higher than the patient's heart, the pressure
exerted by the fluid column will "run away from the transducer",
falsely lowering the pressure.
|
| 13. |
Zeroing
the Transducer
- Remove
the cap on the vent port (stopcock above the transducer).
- Ensure
that the vent port is level with the 4th intercostal space,
midaxillary line).
- Open
the stopcock above the transducer to air (turn it "OFF" to
the patient line).
- Note:
the cap should be removed before turning the stopcock.
- Select
the pressure sensor on the monitor and select "zero".
- When
completed, the monitor will display a message that indicates
"zero completed, offset is x mmHg". The offset should
be between +32 and - 32 mmHg. If it is outside that
range, a new transducer should be obtained.
|
13. |
- During
zeroing, the system is closed only to the patient's pressure.
All other sources of pressure (which cannot be removed) remain
exerted against the transducer (e.g. atmospheric, flush device).
The monitor is then told to call this pressure "zero".
- When
the transducer is reopened to the patient pressure, the pressure
reading will reflect only that of the patient. (This is the
same principle used when zeroing a bathroom scale; one ensures
that the scale is at "zero" before adding your own weight).
- If
the dead-end cap is left on, or the stopcock incorrectly turned,
the monitor will be unable to find an uninterrupted path to
locate "zero" and will display an error message. Recheck
the stopcock and rezero if this happens.
|
|
14. |
When
the monitor indicates zeroing is complete, turn the stopcock
closed to the vent port (air elimination port) first.
- Once
the port is closed to the air, reapply the dead-end cap.
- When zeroing completed remove non-sterile gloves and perform hand hygiene.
|
14. |
- If
the cap is reapplied when still opened to the vent port (air
elimination port), pressure is transmitted back to the transducer
as it hits a "solid wall" from the cap.
- The
dead-end cap protects against air entry or "bleeding out".
- In accordance with the MoHLTC 4 moments ogf hand hygiene and LHSC infection control policies in an effort to reduce risk of transmission of microorganisms and secretions.
|
| 15. |
You
should now identify a pressure waveform and pressure recordings. |
15. |
|