|
Instrumentation
and Technology
High
Resolution Sector Field Inductively Coupled Plasma Mass Spectrometry*
*HR-SF-ICP-MS*
HR-SF-ICP-MS is The Gold
Standard for analysis of Trace Elements in Biological Matrices.
- Interference free
measurement of almost the whole periodic table and in almost all
matrices
- From ultra-trace to
matrix components
- Quantitative, Isotopic
and Species Information
The Trace Elements Laboratory, at LHSC uses both the ELEMENT 1 (previously
Finnigan MAT), and the Thermo Scientific ELEMENT 2 (both HR-SF-ICP-MS).
They outperform Low Resolution "Quadrupole" ICP-MS, in
detection limits, sensitivity, interference elimination, and dynamic
range (see below).
These 2 Instruments are both double focusing magnetic sector field
ICP-Mass Spectrometers.
Both perform identically, using exactly the same technical principles.
Click
here to watch an ion flight demonstration movie
ICP-MS is the accepted and most powerful technique for the analysis
and quantification of trace elements. Its applications range from
the semiconductor industry to geological and environmental analyses,
from biological research to material sciences. The most severe limitation
of ICP-MS is polyatomic interferences on the elemental signals,
originating from Argon and/or the sample matrix.. Elimination of
interferences enables accurate and reliable quantitative multi-element
analyses at trace levels, even without sample preparation.
The ELEMENT redefines
high resolution ICP-MS by setting a new standard for productivity
at the highest level of analytical performance. With the experience
of more than 500 installed ELEMENT's, Thermo Fisher Scientific,
the leader in high resolution ICP-MS, has rigorously optimized the
ELEMENT for ease-of-use, stability and productivity. Autotuning
of all parameters, including lenses, gas flows and the torch position
ensures a reproducible and reliable system setup. Refined ion optics
deliver high transmission at high resolution. The Thermo Scientific
ELEMENT is the high resolution ICP-MS for routine productivity in
your laboratory.
With the ELEMENT there
is always access to the most powerful technique to solve a specific
problem, combined with the highest sensitivity and the best signal-to-noise
ratio. If there is a solution using
ICP-MS, the ELEMENT has it. No other ICP-MS system can make this
claim: ultimate sensitivity at all resolution settings, ultimate
selectivity through high mass resolution, 'hot' and 'cold' plasma
conditions, high speed for transient signals, high stability for
isotope ratios, full automation for routine productivity and full
access to all instrumental parameters for method development.
- Multi-element analysis
across the periodic table covering a mg/L to sub pg/L concentration
range - Compatible with inorganic and organic solution matrices
and solids
- High mass resolution
to access spectrally interfered isotopes - Produces unambiguous
elemental spectra
- A multi-elemental
detector for transient signals - For example, CE, HPLC, GC, FFF
and laser ablation
- High precision isotope
ratios - On non-interfered or interfered isotopes
- Fully automated tuning
and analysis - In conjunction with a comprehensive, customizable
quality control system
- Reliability and robustness
to serve as a 24/7 production control tool - Highest sample throughput
- Highest flexibility
and accessibility to serve as an advanced research tool
HR-SF-ICP-MS*
- Argon gas passes through
a quartz tube (Torch).
- The torch is surrounded
at the end by a copper or silver coil (Load Coil), connected to
a radio frequency generator (RF).
- As power is applied
to the load coil from the RF, electrical and magnetic fields are
established at the end of the torch.
- When a spark is applied
to the argon flowing through the torch, electrons are stripped
off the argon atoms, forming positive argon ions.
- These ions collide
with other argon atoms, forming an argon discharge or Plasma (about
6000°C).
Sample Introduction
- Diluted samples are
introduced through a peristaltic pump to a nebuliser, which produces
a very fine aerosol within a spray chamber.
- The aerosol is carried
by the argon gas into the torch, and then to the high temperature
plasma, where the elements in the sample transported, are desolvated
and vapourised.
- Dissociation occurs
and most elements are fully ionized.
Plasma Interface and
Transfer Optics
- Ions are then brought
into the Mass Spectrometer via Interface Cones
metal disks
with tiny apertures
- These apertures convert
the ions into an ion beam, which is then focused using an electrostatic
ion lens, under high vacuum.
- Photons are blocked
from entering the interface.
- Once the ions enter
the mass spectrometer, they are separated by their mass to charge
ratio.
- Quadrupole instruments
go no further than this before ion detection.
- Consequently they
cannot eliminate mass overlap.
Double Focusing
- High Resolution instruments
allow elimination or reduction of most interferences (Spectral
and Polyatomic) by using both a Magnetic Sector Magnet and an
Electric Sector Chamber.
- The Magnetic Sector
disperses both ion energy and mass, and focuses all the ions with
diverging angles of motion coming from the entrance slit, past
the ion lens.
- The Electric Sector
disperses only ion energy, and focuses the ions onto the exit
slit, just before the detector (SEM).
- This arrangement is
called Double Focusing.
- See internal view
of instrument below
Resolution
- The Resolution can
be changed by adjusting the width of the entrance and exit slits
into the Spectrometer.
- These are called Low,
Medium, and High Resolution.
- Some elements have
overlapping interference masses, and require a narrower slit,
thereby separating or resolving the interferences.
- ICP-MS (Inductively
Coupled Plasma Mass Spectrometer) analysers have long been recognized
as the most powerful technique for the analysis of trace elements.
- The most severe limitation
with quadropole and collision-cell ICP-MS are the polyatomic interferences
on the element signals, originating from the sample matrix and
Argon.
- High Mass Resolution
is the Gold Standard for the identification and elimination of
these interferences.
- The resolved element
ions are then detected or counted by a detector, which counts
the number of ions, and amplifies them.
- The more ions of any
particular element hitting the detector correspond to a higher
concentration of element in that sample.
- By using Standard
solutions and Control solutions, concentrations of unknown samples
can be determined accurately.
ICP-MS can also
simultaneously determine up to 40 or more elements in 1 single
sample.
Advantages of using High Resolution ICP-MS:
- Highest level of analytical
performance
- Multi-element analysis
across the periodic table covering mg/L to pg/L
- Variable Resolution:
The resolving power from 300 for Low Resolution, 4000 for Medium
Resolution and 10,000 for High Resolution allows the choice of
resolution for each element to be analysed without any interferences
and without over-resolving it.
Shown below is an example of the separation of Chromium in urine
from the
interferences of 40Ar12C and 35Cl16O1H
in Medium Resolution.
|
|
|
Qudrupole
Resolution
|
Low
Resolution
|
Medium
Resolution
|
|
|
|
|
|
|
Shown below is an example of that using the element Arsenic at m/z
75 separated
in High Resolution from the interference 40Ar35Cl.
|
|
|
Quadrupole
|
Low
Resolution
|
Medium
Resolution
|
High
Resolution
|
|
|
|
|
|
|
|
- Do not have to apply
ratio calculations due to interferences
- Ultimate sensitivity
at all resolution settings, at Low Resolution Indium at 1 ppb
will give a count greater than 1000,000 counts per second
- Ultimate selectivity
through high mass resolution
- The combination of
magnetic and electrostatic fields results in the double focusing
high resolution properties of the Element. A magnetic field regulator
with a high power stage controls the changing of the magnetic
field. This results in the fastest scan speed possible using a
magnet sector.
- Dual detection system.
The element is equipped with a discrete dynode detector. The detector
is linear from ppq to ppm. This allows for analysis of many elements
in one single analysis. The Dynamic range is > 109 linear with
automatic gain calibration.
- Use of Guard Electrode
on torch decreases the ion spread, resulting in an increase in
the ion transmission and much improved sensitivity.
- Guaranteed off-peak
background noise of < 0.2 cps for all three resolutions, this
allows for achieving the lowest detection limits
- Excellent signal stability:
- <1% RSD over
10 minutes.
- <2% RSD over
1 hour.
For further
information see the ELEMENT2
Brochure by Thermo Scientific
Acknowledgements
Thanks to
Thermo Fisher Scientific for the above documents.
|