|
OBTAINING A CSF SAMPLE FROM AN
INTRAVENTRICULAR CATHETER
IN CCTC |
- Obtain Equpment
- Prepare to Obtain Sample
- Obtain Sample
- Transfer Sample
|
- Restore Circuit
- Label Sample
- Document
|
| PROCEDURE |
RATIONALE
FOR PROCEDURE |
| 1. |
Obtain Equipment
- Face mask with shield
- Sterile gloves
- Sterile towel
- Chlorhexidine 2% with 70% alcohol swabs
- 3 ml syringe with needleless cannula
- Sterile specimen container (with orange lid)
|
1. |
Samples should be represent a freshly collection. Bacterial counts may multiply within drainage bag collections and provide inaccurate culture results. |
|
2. |
Prepare to Obtain Sample
- Perform hand hygiene.
- Apply mask, don sterile gloves.
- Identify sampling port (port closest to patient catheter).
- Turn 3-way stopcock off to pressure tubing (pronged side of stopcock toward the pressure transducer).
- Close slide clamp distal to sampling port
- Open sterile towel on bed and position sampling port on top of towel.
- Scrub port vigorously with chlorhexidine/alcohol swab.
- Allow cleaning solution to dry for 1 minute.
|
2. |
Aseptic technique should be maintained. CSF contains glucose, supporting bacterial growth.
With a 3-way stopcock, flat sides represent open pathways . The pronged side of the stopcock is off.

To draw a sample of CSF, the circuit must be open between the catheter and sampling port.
Cleaning solution must dry on surface to maximize antimicrobial properties. |
| 3. |
Obtain Sample
- Slowly aspirate 2-3 ml of fluid into syringe over 1-2 minutes. Do not force or generate excessive vacuum.

|
3. |
Slow aspiration will allow gentle removal of sample without generating excessively negative pressures. |
|
4. |
Transfer Sample
- Empty contents of syringe into sterile collection chamber.
- Tightly seal lid.
|
4. |
The lab will not accept specimens in syringes due to risk for spillage/exposure of body fluids. |
5. |
Restore Circuit
- Open slide clamps.
- Return stopcock to presampling position as per written physician orders.
- Check all stopcocks and connections.
- Reconfirm level.
- Remove non-sterile gloves and perform hand hygiene, remove faskmask and shield.
|
5. |
To ensure ICP drainage system is returned to therapeutic position. |
| 6. |
Label Specimens
NOTE: ALL CSF CULTURE SAMPLES SHOULD BE ORDERED "STAT" AND A GRAM STAIN REQUESTED
- Place labels on specimens.
- Verify that the label name and patient
are correct.
- Sign sample requisition and record time
sample was drawn.
- Place sample in plastic biohazardous bag with label in outside pocket.
- Remove non-sterile gloves and perform hand hygiene.
|
6. |
Samples will be discarded by lab if unsigned. |
| 7. |
Document
- List sample collected on the lab collection section of nursing
flow sheet.
- Document on Kardex.
|
7. |
Kardex communicates to subsequent shifts to monitor for culture results. |