|
PRISMAFLEX(TM)
Introduction
Differences
|
|
INTRODUCTION
The
PRISMAFLEX(TM) is the newest generation of technology for CRRT (hemodialysis)
offered by Gambro.
The
following overview has been developed to provide a brief introduction
to the PRISMAFLEX(TM) machine. This information is intended to introduce
participants to the main differences between the current technology
and the PRISMAFLEX.
PRISMAFLEX
educational sessions will provide more detailed instruction regarding
the technology.
How
the Prismaflex(TM) is different from Prisma(TM)
(major differences):
- The Prismaflex(TM)
has an additional pump that allows the delivery of fluid
into the blood pre filter. This circuit is called the Pre Blood
Pump (PBP). Fluid delivered into this circuit will be accounted
for in the calculation of net fluid balance. In other words, if
200 ml is being delivered via the PBP and there is 100 ml of fluid
removal, the machine will automatically remove the extra 200 ml.
Examples of solutions that can be delivered via the PBP include
anticoagulants (heparin or citrate).
Prisma(TM) does not have a PBP. Infusions
delivered prefilter are currently administered via an external
infusion pump and are not accounted for in the net fluid removed
as displayed by the machine. Currently, when infusions are administered
(e.g. citrate), the fluid removal must be increased to match the
infusion rate to provide a true net balance. Adjustments for the
citrate infusion will no longer be required with the PRISMAFLEX(TM).
When running citrate, the fluid balance will still
need to be increased according to the systemic calcium chloride
infusion.
- The tubing
for the Prismaflex(TM) is larger. This
will lowers the resistance to flow and can reduce the amount of
pull required to remove blood from the patient. This can cause
access pressures to become less negative. Occassionally, access
pressures may become low enough to trigger access disconnect alarms
(access disconnect alarms occur when the access pressures are
higher than <10).
The solution to erroneous access disconnect alarms is to increase
the blood flow rate until the access pressure again falls below
<10 (it take more "pull" to move blood through the
filter at a higher rate).
- The Prismaflex(TM)
has much higher maximum flow rate capabilities for blood
flows, dialysate and replacement rates. The Prisma(TM)
can only achieve blood flow rates of 150-180 ml/min (depending
upon model). The current combined flow rate for dialysate PLUS
replacement volumes is 4 L/hr. The Prismaflex(TM)
can deliver up to 8 L/hr per hour (dialysate, replacement
plus PBP). To achieve flow rates this high, a larger filter with
higher blood flow capacity needs to be used. We will continue
to use the same type of filter with similar flow rates to our
current practice.
- Currently,
Prisma(TM) is initially set-up in CVVHDF
mode. This provides the flexibility of being able to change the
therapy to CVVH or SCUF, without changing the filter. If the Prisma(TM)
is set-up as CVVH, dialysis can only be introduced by discontinuing
the current treatment and setting up a new filter. While CVVHDF
mode set-up provides the flexibility to adjust the treatment downward,
it does limit the hemofiltration maximum flow rate to 2 L/hr.
Higher hemofiltration rates can be run if the circuit is set-up
initially as CVVH, however, dialysate cannot be added without
starting a new filter. Prismaflex(TM) is
a much more flexible system that allows any therapy change.
The same Prismaflex(TM) filter can be run
as predilution, postdilution, or a combination of both, when hemofiltration
therapy is used. Currently, Prisma(TM)
can only be run as predilution OR postdilution, depending
upon the type of filter being selected. Currently, we only stock
predilution sets for the Prisma(TM).We
will continue using the predilution filters at this time.
- An air elimination
port similar to those found on intermittent hemodialysis circuits
is incorporated into the Prismaflex(TM) circuit.
This is not a component of the Prisma(TM).
Assessment of the air-fluid level will be required hourly and
prn. A filter to prevent blood from entering the machine is also
used.
- Technical
differences exist among the filters used, methods for priming,
heater and heater tubing and method for bag changes. This will
be demonstrated in class.
- Scales are
still used to measure weight changes, however, the new Prismaflex™
is less susceptible to nuisance alarms. Scales are now located
at waist level, with an improved ergonomic design for bag changes.
- Data
storage capabilities are available with the Prismaflex(TM).
References:
Gambro Training Manual 1 and 2
Slides from Gambro Training package, reproduced with permission
Last UpdateMarch 30, 2010
|