|
PROCEDURE
FOR DRAWING BLOOD FROM
AN INDWELLING LINE IN CCTC
|
- Select Patient
- Obtain Equipment
- Prepare Equipment
- Prepare Environment
- Prepare Access
Cap
- Connect Blood Withdrawal
System
|
- Position Stopcock
for Sampling
- Collect Discard
Sample
- Collect Samples
- Flush Catheter
- Flush Sampling
Port
- Label Specimens
- Document
|
| PROCEDURE |
RATIONALE
FOR PROCEDURE |
| 1. |
Select Patient
- Verify order
is written or blood test is supported by preprinted order/protocol.
- Order laboratory tests in PowerChart
and obtain labels.
- Check labels with patient's armband at bedside to verify correct patient is having blood drawn.
DO NOT REMOVE
CLEARLINK CAP TO DRAW SAMPLES. BLOOD MUST BE DRAWN THROUGH
THE NEEDLELESS CLEARLINK ACCESS CAP.
|
1. |
A medical order or supportive protocol
is required by the College of Nurses.
The procedure for blood drawing maintains
a closed system. |
|
2. |
Obtain Equipment
- Obtain non-sterile gloves and asess risk for need of a face shield.
- LHSC Routine Practices Policy
- LHSC Hand Hygiene Policy
- Vacutainer with indwelling needle for blood collection
- 2% chlorhexidine and 70% alcohol swab
- Blood collection tubes as required to obtain ordered lab tests
- Blood gas syringe if required
- Two additional RED TOP blood collection tubes. One will be
used to collect the discard sample and the other will be used
to collect backflush.
- Orde labels from powerchart
Note: If you are not collecting any blood into
vacuum tubes (i.e, blood gases, glucometer samples or point-of-care
testing), a syringe can be used to collect the discard sample.
A sleeve with slip lock and needlesless access cannula is still
required to collect the backflush into a red top tube after blood
is collected by syringe. |
2. |
The needless system reduces the risk for
stopcock contamination and needlestick injuries.
The Lever-Lock(TM) cannula with the clip
is used to prevent disconnection during blood withdrawal.
Collection of backflush solution into a disposable blood tube
reduces exposure to blood products.
|
| 3. |
Prepare Equipment
- Assemble vacuntainer, tubes and/or blood gas syringe
|
3. |
|
|
4. |
Prepare Environment
- Explain procedure to patient/family.
- Press the alarm silence button and hold
for 4 seconds if samples are to be drawn from an arterial line.
|
4. |
Blood sampling from an arterial line reduces patient discomfort
from needle sticks, however, patients should be advised that they
may feel a warm sensation in their extremity during line fllushing.
When the silence button is pressed once, all alarms are suspended
for 2 minutes. When the button is depressed for 4 seconds, the
alarm is suspended for 5 minutes. This will prevent nuisance alarms
during blood sampling from the arterial line. |
|
5. |
Prepare Access Cap
- Perform hand hygiene and don non-sterile gloves, face shield if required.
- Scrub the needleless access cap with
2% chlorhexidine and 70% alcohol swab
- Allow chlorhexidine to dry for 1 full minute before connecting
access cannula

|
5 |
Scrubbing action loosens surface pathogens.
Follow routine precautions and the 4 moments of hand hygiene outlined by LHSC and the MoHLTC.
Antimicrobial effects begin after the chlorhexidine
has dried. Evidence favours the use of chlorhexidine for skin
preps, to reduce blood stream injections. There is currently no
evidence to compare chlorhexidine for injection port cleaning.
The Centre for Disease Control (CDC) recommends 70% alcohol or an or an iodophor for cleansing injection
sites. Chlorhexidine adheres to provide prolonged gm positive
antimicrobial properties. |
| 6. |
Connect Blood Drawing System
Connect vactainer with indwelling needle to the Clearlink cap on stopcock.
|
6. |
This provides blood access. |
| 7. |
Position Stopcock for Sampling
Open stopcock toward patient and sampling port by turning the
white prong of the 3-way stopcock toward the flush device. Turn
stopcock to 90 degrees.
|
7. |
This position is open to the patient and the sampling port and
closed to the flush device. |
| 8. |
Collect
Discard Sample
- Insert one of the red top tubes
into the vaccutainer with indwelling needle.
- Depress the blood tube to activate the
vacuum.
- Turn the stopcock off toward the flush
system.
- Obtain a
MINIMUM 3 ml discard if from an arterial line (including
if glucometer and blood gas sampling).
- An 8 ml discard sample is required when
drawing blood from a central line.
- Dispose of discard samples into the sharps container.
Note: If you are not collecting any blood tubes,
you can collect discard sample into an extra syringe.

|
8. |
An adequate discard sample is even more
important when a very small blood volume is withdrawn, as the
potential for signficant sample dilution is greater.
LHSC Sharps Handling Policy |
| 9. |
Collect Blood Samples
- Turn the white stopcock prong into a
45 degree position before releasing the vaccum or changing tubes/syringe.
- Remove the discard sample and connect
the first blood sample tube or syringe.
- ALWAYS turn stopcock to
45 degrees before releasing the vaccum on the
blood tube, before changing tubes.
- Draw blue top tubes for aPTT
last, but BEFORE blood gas sample.
 |
9. |
When the stopcock is turned to 45 degrees, the stopcock is off
in all directions.
If tubes or syringe changes are made when the stopcock is open
to the sampling port, blood splatter may occur.
If the stopcock is turned off to the sampling port in a 90 degree
position, heparinized flush solution can enter the next sample
drawn, diluting or contaminating the results. |
| 10. |
Flush
the Patient Catheter
- When all of the desired specimens have
been collected, turn the white prong on the stopcock toward
the sampling port.
- Pull the flush device located below
the transducer to activate the fast flush mechanism.
- Flush until there is no visible evidence of blood.
- Flush central lines or PA lines for one full minute.

|
10. |
The prong points toward the "off"
position. This position is open to the patient and the flush device
but closed to the sampling port.
Any blood remaining in the line may trigger
clot formation. Central lines and PA catheters have long internal
lumens that require a longer flushing time for adequate clearance. |
| 11. |
Flush the Stopcock
- Connect
the second extra syringe or blood tube to the sampling port.
- Turn the stopcock so that the white
prong points toward the patient's catheter.
- Activate the flush device and back flush
saline into the red tope tube or syringe.
- Manipulate the lever lock to ensure
all blood is cleared from the Clearlink cap.
- Replace Clearlink caps that have residual blood.
- Remove non-sterile gloves, perform hand hygiene.
|
11. |
Blood provides a medium for bacterial growth.
Backflushing into a contained blood tube
reduces caregiver exposure to blood.
In accordance with the MoHLTC 4 moments of hand hygiene and LHSC infection control policies in an effort to reduce risk of transmission of microorganisms and secretions.
LHSC Hand Hygiene Policy
LHSC Routine Practices Policy
|
| 12. |
Label Specimens
- Place labels on specimens.
- Verify that the label name and patient
are correct.
- Sign sample requisition and record time
sample was drawn.
- Send blood samples to the lab in a biohazardous bag.
|
12. |
Samples will be discarded by lab if unsigned. |
| 13. |
Document
- List sample collected on the nursing
graphic record.
- Ensure an order has been written to
obtain blood sample.
- If blood sample has been drawn according
to a prn protocol, document in the AI record the reason the
sample was obtained.
- Record bloodwork results on lab flowsheet and review results with physician.
|
12. |
Medical orders are required for blood sampling.
PRN protocols require documentation of reason sample was collected.
Note that a written order is required for
non-protocol driven samples (e.g., cultures, cardiac or hepatic
enzymes, urea and creatinine).
To identify need for intervention. |