|
PULMONARY ARTERY CATHETERS (PA):
STANDARDS OF NURSING CARE IN CCTC (SONC)
|
- Monitor ECG
- Provide Emergency
Equipment
- Set-Up Hemodynamic
Circuit
- Maintain Accuracy
- Display Waveforms
- Monitor Cardiovascular
Function
- Maintain Closed System
- Obtain Chest Xray
|
- Select Appropriate
Ports for Infusions
- Obtain Blood Samples
- Change Dressings
- Reposition PA Catheter
- Analyze CVP and
PWP Waveform
- Measure Cardiac
Output
- Document Placement
- Paceport (TM) Catheters
|
| STANDARD
OF NURSING CARE |
RATIONALE
FOR STANDARD |
|
1. |
Monitor ECG
All CCTC patients with PA
catheters will have continuous ECG monitoring. See Standard
for ECG Monitoring.
|
1. |
To provide continuous monitoring with prompt detection of changes
in heart rhythm. PA catheters can slip into the right ventricle
and produce ventricular ectope.
|
| 2. |
Provide Emergency Equipment
Access to ACLS equipment (e.g., defibrillator,
drugs) is required for all patients with a PA catheter.
|
2. |
Migration of the right heart catheter into the right ventricle can
cause life threatening arrhythmias.
|
| 3. |
Set-Up
Hemodynamic Circuit
Clinical Nurses in CCTC are responsible
for the priming, zeroing, leveling, and maintenance of hemodynamic
pressure monitoring circuits and for the assessment and monitoring
of hemodynamic pressures and waveforms.
Clinical Nurses may flush hemodynamic monitoring
circuits as required to maintain patency
Exception:
left atrial lines and intra-aortic balloon
pump lumens are maintained with pressurized heparinized saline BUT
are not manually flushed by Clinical Nurses.
|
3. |
Clinical Nurses in CCTC have received additional education in the
care of invasive monitoring circuits. Refer to Procedure for Hemodynamic
Monitoring Circuits or Removal
of a Right Heart Catheter.
See also: Standard for Care
of a Patient on an IABP
|
| 4. |
Maintain Accuracy
Hemodynamic transducers are zeroed at each initial setup, with
the air-fluid interface (stopcock above transducer) leveled to the
mid-axillary line.
Transducer levels should be validated at
the beginning of each shift, prior to each PWP or CVP measurement,
following position changes, and prn to validate hemodynamic pressures.
Document level confirmation in the graphic
record. Documentation is required at the start of each shift and
q6h, prior to each CVP and PWP measurement, and following transducer
repositioning.
|
4. |
Zeroing eliminates the effect of atmospheric pressure on measured
hemodyamic values.
Leveling eliminates the influence
of hydrostatic pressure on the transducer. A transducer that is
positioned below the patient's heart will produce falsely elevated
pressures and a transducer positioned above the patient's heart
will produce falsely low pressures. |
| 5. |
Display Waveforms
The tip of the PA catheter must be connected
to a monitor that provides a continuous waveform display. Continuous
monitoring with waveform display is required during transport. |
5. |
The waveform from the tip of the right
heart catheter must be displayed continually to demonstrate the
catheter tip location within the heart. Continuous monitoring is
required to promptly detect catheter migration into the right ventricle
or pulmonary wedge position.
|
| 6. |
Monitor Cardiovascular Function
Monitor BP (Systolic/Diastolic/Mean), PAP
(Systolic/ Diastolic/ Mean), HR, RR and SpO2 continuously and document
q1h in the graphic record, for all new admissions, and for patients
with arterial lines or pulmonary artery catheters, who are receiving
medications that influence blood pressure, or whose neurological,
cardiovascular or respiratory status has varied during the previous
24 hours.
Nurses may decrease the frequency of BP monitoring
in patients who remain stable without intervention, and who do not
meet any of the requirements noted above. Monitoring frequency must
be increased if the patient's condition deteriorates. When the frequency
of monitoring is decreased, the rationale used to support the frequency
change must be documented in the AI record.
|
6. |
To ensure prompt identification and documentation of patient status.
See also Standards for Respiratory
and Neurological Monitoring. |
| 7. |
Maintain
Closed System
All stopcocks must have dead-end (non-vented) luer lock caps or
luer lock connected infusions. This includes stopcocks located on
transducers.
Hemodynamic circuits are changed with
each new line and prn.
|
7. |
To prevent accidental entry of air or contaminants. Accidental movement
of a stopcock that does not have a dead end luer lock cap can result
in hemorrhage/air embolus.
|
| 8. |
Obtain
Chest Xray
Obtain chest xray daily, following all central
venous or right heart catheter insertions, or if the right heart
catheter is advanced > 5 cm.
|
8. |
To identify complications including: pneumothorax, hemothorax, hydrothorax,
catheter kinking, catheter placement (e.g. right atrial placement
of CVP catheters or distal placement of right heart catheter), pulmonary
infarction or hemorrhage.
Chest xrays do not need to be
repeated for simple catheter withdrawals, unless there is ongoing
concern regarding placement.
Daily and post-advancement xrays are done to assess for spontaneous
catheter migration and to rule out distal placement, or to monitor
for complications (e.g., hemorrhage, infarction, pneumothorax).
Distal placement may be detected on xray, in the absence of a spontaneous
wedge tracing.
|
| 9. |
Select Appropriate Ports for Infusions
Continuous infusions that affect BP or cardiac
output should not be administered via the blue proximal injectate
port of the PA catheter.
If the proximal injectate port is the only site available for the
administration of vasoactive drugs, utilize the proximal port but
do not perform cardiac output measurements.
Although blood products can be infused through
any central line, it is preferable that the blue injection port
of the right heart catheter be avoided. |
9. |
During cardiac output measurements, 10 ml of fluid is administered
into the proximal injectate port. This would result in the administration
of a bolus of vasoactive drug. While performing the cardiac output
injections, the infusion would be disrupted. After completion of
the cardiac output measurement, a delay would occur before the infusion
would reach the patient.
Administration of a vasoactive drug via a peripheral vessel places
the limb at risk for injury. This risk outweighs the potential benefit
associated with PA catheter monitoring.
Blood products may reduce patency
of this lumen making measurement of cardiac outputs impossible.
|
| 10. |
Obtain Blood Samples
Clinical Nurses may draw blood from the pulmonary artery catheter
using a stopcock. |
10. |
See Procedure for
Blood Withdrawal and Blood
Gases Venous. |
| 11. |
Change Dressings
Central line dressings are changed q 2 days and prn if occlusivity
is disrupted. Document date of change in the graphic record and
kardex. Record any redness or abnormal findings in the AI record
and report to the physician. See Procedure
for Arterial and Central Line Dressings.
Lines are changed prn for evidence of site
infection or sepsis. |
11. |
Dressings are change q 2 days in critically ill patients to allow
site inspection and to reduce bacterial colonization (CDC recommendation).
Skin condition, diaphoresis and site oozing necessitates more frequent
dressing changes in critically ill patients.
For stable, long stay patients with
central lines, Opsite 3000 (TM) dressings
can be changed q7days as per general hospital policy, as long as
occlusivity is maintained.
The Centre for Disease Control
does not recommend routine line changes.
|
| 12. |
Reposition PA Catheter
Clinical Nurses may not insert or advance a right heart catheter.
Clinical Nurses in CCTC may partially withdraw a right heart catheter
from a persistent PWP or RV position without a physician's order.
Clinical Nurses may remove a PA catheter with a physician's order.
Prior to withdrawal of a PA catheter tip from an RV to RA placement,
switch proximal infusions to the introducer site.
|
12. |
Currently, the College of Nurses does not permit advancement of
a right heart catheter. Clinical Nurses in CCTC have received additional
education in the care of right heart catheters. Refer to Procedure
for Partial Withdrawal or
Removal of a right heart
catheter.
During withdrawal of the PA catheter, or manipulation by the resident,
the proximal ports (right atrial, right ventricular ports) may infuse
into the introducer or the sleeve. This can cause fluid or blood
to backup into the sleeve, with disruption of drug delivery.
|
|
13. |
Analyze CVP and PWP Waveform
CVP and PWP measurements are obtained from
a waveform printout. Waveforms are printed and analyzed to identify
the right atrial pressure at end-expiration and the base of the
"V" wave.
Waveform analysis is identified on a paper recording of each right
atrial and PWP tracing. Each waveform is posted in the clinical
record.
|
13. |
The goal for right atrial pressure and PWP monitoring is to identify
the pressures during end-expiration and end-diastole. Although the
pre-C wave best reflects the end-diastolic pressure, it is often
difficult to identify, reducing inter-rater reliability. For consistency,
we measure the pressure during mid-diastole (during diastasis) because
it is easy to find and reflects a pressure that is similar to the
pressure at end-diastole.
|
| 14. |
Measure Cardiac Output
Measure cardiac
output q 6 h and prn following changes in vasoactive drugs, fluid
administration or to reassess changes in the patients hemodynamic
status.
Measure arterial and venous gases with each
cardiac output, and enter data into the Critbase program. Measure
hemoglobin OD and prn if indicated (e.g., to assess potential bleeding,
following blood transfusion), and utilize most recent value to determine
oxygen content variables.
Measure CVP and PWP prior to cardiac output
measurement. Obtain a minimum of 3 injections, editing measurements
to average 3 samples that have a variance of +/-10%.
|
14. |
See Procedure for Measuring
Cardiac Output.
See Procedure for
Blood Withdrawal; Blood
Gases Arterial.and Blood
Gases Venous. |
| 15. |
Document
At the start of each shift, identify PA catheter
placement and record the distance marker at the entrance point to
the introducer hub in the intravascular device section of the flow
sheet. Each single black line represents 10 cm, while each
thick line represents 50 cm. The markings at the insertion site
represents the amount of catheter "in" the patient.
Assess catheter patency and document in the
intravascular devices section of the flow sheet at the start of
each shift and q 4 h.
Document catheter repositioning procedures
in the AI record and record changes to the distance marking in the
intravascular device section of the flow sheet.
Monitor waveform continuously. Determine CVP and PWP q 6 h and prn
as per #14. Document pressures in the graphic record.
Document intravenous infusion in the 24 hour
assessment record and sign for all infusions in the MAR, including
heparinized flush solution.
Record IV infusion volumes in the fluid balance
record.
Record vasoactive drug doses in the graphic record section of the
flow sheet.
Document dressing changes in flow sheet. Star
and DAR abnormal findings.
|
15. |
To assist in recognizing when catheter migration has occurred and
to ensure distal placements are recognized and corrected.
Waveforms identify the location of the catheter tip.
The catheter tip can spontaneously retract into the right ventricle
and cause ventricular ectope, or advance into the distal pulmonary
circulation and compromise blood flow. Catheter tip movement can
occur without a change in the insertion point markings. |
| 16. |
Paceport (TM) Catheters
If a Paceport(TM)
catheter is being used to pace a patient, the right ventricular
side arm patency must be maintained as per any central line. Luer
lock connections must be used, and it is recommended that an infusion
(verses heparin lock) be used to maintain line patency. A flow rate
of 1-2 ml is adequate by volumetric pump. Inotropes and vasoactive
drugs may be infused through this lumen during pacing if needed.
Following insertion of a pacemaker wire,
wedging should be performed cautiously, observing for any loss of
capture of the pacemaker. If capture issues occur, do not use the
catheter to perform PWP measurements during pacing.
Cardiac output measurements may be performed
as long as capture issues do not occur during injection.
|
16. |
The RV is an excellent location to introduce drugs that require
central administration due to the large volume of blood for rapid
dilution.
The higher RV pressure increases
the potential for line thrombosis. Volumetric pumps at a low flow
rate are usually sufficient to maintain the positive pressure that
prevents line blockage.
During wedging, the right heart
catheter migrates forward at the tip. In some patients, this can
cause movement of the pacemaker wire. In some patients, positioning
of the right heart catheter to facilitate pacing results in poor
placement for wedging; it may not be possible to perform both functions.
Assessment should be done on a patient-to-patient basis. |