![]() |
About Us | ![]() |
Patients, Families & Visitors | ![]() |
For Health Professionals | ![]() |
Careers | ![]() |
Research & Training | ![]() |
Ways to Give | ![]() |
|---|
PROCEDURE FOR DRAWING MIXED
VENOUS BLOOD GASES
|
|
|
Rationale | ||
|---|---|---|---|
| 1. | Mixed
venous blood gases must be drawn from the pulmonary artery port of the
right heart catheter or a right ventricular lumen
(provided that the right ventricular port is situated in the right ventricle)
Proceed to withdraw a discard sample as per procedure for withdrawal blood samples from an indwelling line. |
1. |
|
| 2. | Ensure that stopcock is turned to 45o between syringe changes. | 2. | Forty five degrees is off in all 3 directions. This will prevent any heparinized saline from entering the tubing between the stopcock and the patient catheter between syringe changes. |
| 3. | Evaluate pulmonary artery tracing to verity that the catheter is not wedged. | 3. | Aspiration from a catheter that is wedged can cause arterialized blood to be drawn into the syringe, falsely elevating the result. |
| 4. |
|
4. | Vigorous aspiration can cause arterialized blood from the pulmonary capillary to be drawn into the syringe, falsely elevating the result. |
| 5. | If
the mixed venous oxygen saturation is > 80%, repeat the sample with careful
attention to technique. If a second sample produces an elevated oxygen
saturation, have resident withdraw the catheter and repeat.
If the oxygen saturation remains elevated, the patient should be evaluated for possible ventricular septal defect or patent ductus arteriosa. |
5. |
|
| Label the syringe and place on ice. Either take sample to the point of care testing machine (GEM) or send sample to the lab on ice in a biohazardous bag . | 6. | Ice stops the cells from metabolizing. If there is a delay before the sample is tested, a sample that is left at room temperature may have a false increase in carbon dioxide, decreased pH or lowered PaO2 due to continued cell metabolism. | |
| 7. | Flush
the catheter, stopcock and clearlink cap of blood and replace cap as per blood drawing
protocol.
The
pulmonary artery lumen requires manual flushing for at least one minute
to clear the long narrow lumen.
|
7. | To
maintain catheter patency and reduce potential for bacterial colonization.
Inadequate
flushing will lead to loss of lumen patency.
|
8. |
Remove non-sterile gloves and perform hand hygiene. | 8. |
In accordance with the MoHLTC 4 moments of hand hygiene and LHSC infection control policies in an effort to reduce risk of transmission of organisms and secretions . |
| 9. | Ensure that lab requisition includes FiO2, PEEP, AC/SIMV, PS, Vt, minute volume and temperature if sending sample to the lab. For point of care testing machine (GEM), FiO2, PEEP, minute volume and temperature are required to be inputed into the GEM. See GEM policy: http://www.lhsc.on.ca/Health_Professionals/CCTC/home/gemPolicy.pdf NB: Note the cardiac output. |
9. |
|
| 10. | Document results in the graphic record and communicate any significant findings to the physician and respiratory therapist. | 10. | To ensure appropriate interventions are made. |
|
November,
1988 Last Update: Mach 23, 2010. Brenda Morgan, Clinical Educator, CCTC |