|
PROCEDURE FOR SETUP AND USE OF CODMANTM
INTRAVENTRICULAR DRAINAGE UNIT IN CCTC |
- System Overview
- Obtain Equipment
- Open Unit
- Prepare Panel
- Prime the Unit
|
- Position the Transducer
- Set Drainage Level
- Zero Transducer
- Connect to Patient
|
| SYSTEM OVERVIEW |
| Nurses
may prime and connect the intraventricular drainage system in CCTC.
All CSF draining catheters (intraventricular; lumbar ICP) must be connected
to the CodmanTM External Drainage System for safety. This setup is used for CSF drainage using either the CodmanTM Microsensor Ventricular Drainage Catheter or the standard fluid filled ventricular drainage system. |
The CodmanTM External Drainage System is a closed
system for drainage and pressure monitoring of intraventricular catheters. An
antimicrobial filter located at the top of the graduated collecting chamber
and bottom of the drainage collecting bag prevents backward flow of fluid or
bacteria. The unit must hang upright at all times. If it is placed flat in a
horizontal position, the filter at the top of the drip chamber can become wet,
preventing the system from draining properly.
The
panel has a "0" reference point that must be positioned at the level of
the catheter (in CCTC, the external auditory meatus is used as the reference point).
The drip chamber buretrol is then adjusted up or down until it is at a
level equal to the maximum desired intracranial pressure. If the
system is open to drainage, CSF will drain as soon as the pressure in the
head exceeds the level set by the height of the drip chamber.
The
CodmanTM drainage
unit uses 3-way stopcocks.
The stopcock can be opened in 3 directions simultaneously. The direction the prong points represents the "off" position.This system may be utilized
for continuous pressure monitoring alone, continuous drainage alone or
continuous drainage and pressure monitoring. To obtain an accurate pressure
measurement, the stopcock should be turned "off" to drainage and "open"
to the patient catheter and pressure monitoring system. If pressure
is measured when the drainage unit is open in all 3 directions, the pressure
measured will be a reflection of the both the patient and drainage bag.
|
 |
The Unit
The panel (left) must be positioned
in an upright position, with the "0" reference point on the panel level
with the patient's catheter. In CCTC, we use the external auditory meatus as the
reference. The unit is designed to accommodate a customized lazer level (kept in CN drawer). This level is positioned to ensure that the laser beam is level with the "0" reference.
The drip chamber is then
adjusted to a level equal to the maximum desired ICP (in cmH20 or mmHg).
The mounting panel stopcock
is at the "0" level. |
 |
Buretrol Close-up (left)
-
Roller to adjust drip
chamber up and down
- Drainage tubing
- Filter
- Level of drainage
(drip point)
- Fluid collection
- Arrow for alignment
of drip chamber with level of desired pressure in cm H20 is the column to
the right (red on panel) and mmHg is the column to the left (black on panel).
|
stopcock codman
|
3-Way Stopcock (left)
The prong is pointing to pathways that are closed. This picture on the left displays the stopcock "open" to the patient, the drainage system and pressure monitoring.
|
PROCEDURE |
RATIONALE
FOR PROCEDURE |
| 1. |
Obtain Equipment
1.
CodmanTM External Drainage System
2.
500 ml bag of 0.9% NaCl
3.
1 single transducer kit for arterial line*
4.
IV pole
5.
Laser level (in CN cupboard)or laser level pen designated for EDS lll located at the CN desk
6. PPE - non-sterile gloves
*Note: If you are setting up the drainage system for use with the Codman Microsensor EVD catheter, continue to prime the system with a single hemodynamic pressure set. If using the Microsensor, do not connect the hemodynamic transducer to pressure monitoring; the Codman Express is the preferred method for monitoring.
The use of an infusion for priming provides a closed system flush in the event that you need to flush blood or tissue out of the tubing DISTAL to the stopcock. Pressure tubing provides an additional safeguard against accidental flushing as it will not run if the flush device was inadvertently left open...it will only flow if the flush device is intentionally pulled, or the saline is pressurized.
If you are using a conventional fluid filled ventricular catheter (not the Microsensor), the transducer on the hemodynamic set will also provide pressure monitoring.
|
1. |
|
|
2. |
Open Unit
Open
package with CodmanTM External
Drainage Unit, being careful not to drop extension pieces out; extension
pieces are loosely connected in package.
|
2. |
The
connections are loosely attached to facilitate gas sterilization. |
| 3. |
Prepare Panel
-
Replace the vented white cap located on the mounting panel stopcock (at the "0" reference mark) with a dead-end luer lock cap (provided in kit).
-
Turn
the mounting panel stopcock with the flat side (off) toward the cap.
Always
leave the mounting panel stopcock in this position; if you wish to turn
the stopcock open or closed to drainage, use the stopcock closest to the
patients head.
-
Perform hand hygiene.
|
3. |
This
opens the system between the drainage tubing and drip chamber in preparation
for flushing.
By
only adjusting one stopcock during use, the risk of accidental closure/opening
of the system is reduced. |
|
4. |
Prime the Unit
-
Hang
a bag of normal saline.
DO NOT ADD HEPARIN
OR PRESSURIZE THE PRIMING SOLUTION.
-
Using
the normal saline, prime the pressure tubing as per setting up a hemodynamic
system.
-
Remove
the extension piece at the distal end of the pressure tubing.
-
Remove
the cap from the stopcock located at the end of the drainage tubing (this will be referred to as the "patient stopcock").
-
Connect the distal
end of the pressure tubing to the stopcock where the cap has been removed. Be careful not to contaminate the connections.
-
Turn
the patient stopcock with flat side (off) towards the patient connection end of the circuit.
-
Pull the flush device until
the all air bubbles have been eliminated and fluid
drips readily into the drip chamber.
-
Turn
the patient stopcock with the flat side (off) towards the drip chamber.
-
Remove
the unvented cap at the tip of the drainage tubing, being careful not to
contaminate the end.
-
Flush the circuit until all air has been expelled between the patient
stopcock and end of drainage tubing.
-
Replace
the unvented cap at the end of tubing and tighten.
-
Turn the patient
stopcock to a 45 degree angle from the tip.
|
4. |
Heparin
is not used due to risk of intracranial bleeding. The system is not
pressurized to avoid introducing fluid into the brain.
Once
primed, the pressure tubing will be used to measure intracranial pressure.
The
extension tubing is an unnecessary
access point with ICP monitoring systems.
The
pressure tubing is connected to the stopcock closest to the patient to
minimize the length of tubing that the pressure wave must travel. CSF
drainage often contains blood and debris; adding the flush system at the
patient stopcock allows the circuit to be flushed between the patient stopcock and the drip chamber as required to maintain patency and pressure transmission.
NURSES ARE NOT APPROVED
TO FLUSH BETWEEN THE PATIENT STOPCOCK AND THE PATIENT CATHETER.
The entire circuit must be primed to facilitate drainage. Be careful to maintain aseptic
technique; CSF contains glucose which promotes bacterial growth that can
cause meningitis.
Tightening
maintains sterility and prevents loss of priming fluid. A 45 degree angle
on any stopcock ensures that it is "off" in all directions. |
5. |
Position the Transducer
-
Securely
tape the transducer to the side of the mounting panel with the air fluid
interface (the air evacuation port of the stopcock above the transducer)
at the same horizontal position as the "0" reference on the mounting panel.
-
Connect
the transducer cable to the bedside monitor.
-
Using
the laser level, ensure that the "0" reference on the panel is level with
the outer aspect of the ear canal.
|
5. |
Both the
transducer and the mounting panel stopcock must be level with the external auditory meatus.
By securing both stopcocks at the same point, the transducer and mounting panel can be simultaneously referenced by adjustment of the mounting panel. |
| 6. |
Set the Drainage Level
-
Obtain
an order for the level of the drainage chamber.
-
Loosen the buretrol adjustment screw and move the drip chamber until the small arrow
is aligned with the desired height.
-
Tighten the screw to secure the buretrol position.
NOTE:
A typical order is "10 cmH20" above the head. The right hand column of
numbers (with red writing) is in cmH20. The column to the left is
in mmHg (orders for drip chamber placement are usually in cmH20 - check
with the physician to confirm the correct level).
In
the example of "10 cmH20", the system is correctly placed if the "0" reference
point is level with the patient's external auditory meatus and the small arrow is
aligned at the 10 cmH20 marking. |
6. |
If
the "0" reference is level with the patient's catheter, a drainage level
of 10 cmH20 above the head provides automatic "venting" of fluid from the
head the moment the intracranial pressure rises above 10 cmH20. If
the "0" reference and the drip chamber are both positioned level with the
catheter, drainage would occur the moment the pressure in the head > 0.
Normal ICP is < 10 mmHg. If the drainage is set at a level below normal ICP, drainage will occur even if the ICP is not elevated.
Weaning of the drainage system can be done by raising the level of the drainage unit. Drainage will only occur if the pressure exceeds this higher pressure. Intermittent drainage may also be initiated, where pressure is monitored and drainage only opened if the pressure exceeds a specific level.
Both
the patient and panel stopcocks must be open between the patient's catheter
and the drainage chamber for drainage to occur. When the system is left continuously open to drain and the drainage chamber is positioned to initiate drainage at a set level, ICP can be maintained at a more consistent
level.
Note
that the bedside monitor measures pressure in mmHg; 10 cmH20 is the same
as 7.4 mmHg. Thus, a properly positioned unit with the drip chamber
at 10 cmH20 will drain if:
-
the
stopcocks are open between the patient's catheter and the drip chamber
-
the
pressure in the head exceeds 10 cmH20 (or exceeds 7.4 mmHg on the bedside
monitor).
-
the
catheter/circuit is patent (e.g. free of clot or debris)
-
there
is fluid in the ventricle available to drain (if the pressure is high due
to brain swelling but the ventricles are collapsed, there won't be any
CSF available to drain)
If the circuit fails to drain despite a pressure on the monitor that is > than the level of the drainage chamber, patency can be assessed by lowering the entire drainage unit to below the head.
Notify Neurosurgery if the unit fails to drain in the presence of an elevated pressure despite troubleshooting efforts. |
7. |
Zero the Transducer
-
Select "ICP" as the monitor label.
-
Open the transducer stopcock to air (taped to the mounting panel).
-
Press "zero" on the bedside monitor.
-
If the ICP is > 20, the scale will need to be increased.
-
Perform hand hygiene and don non-sterile gloves.
|
7. |
Zeroing and leveling the transducer ensures the bedside monitors reads the ICP accurately.
The ICP default scale is 20. |
8. |
Connect to Patient
-
Following insertion of the intraventricular catheter, the physician must insert the small white connector into the silicone intraventricular catheter.
-
The connector has a "nipple" end that is inserted inside the catheter tip. This nipple has a luer lock connection for attachment to the CodmanTM unit.
-
Connectors are provided in both the intraventricular catheter kit and the CodmanTM.
|
8. |
|