|
ARTERIAL LINE MONITORING:
STANDARDS OF NURSING CARE IN CCTC (SONC)
|
- Set-Up Hemodynamic
Circuit
- Maintain Accuracy
- Monitor Blood Pressure
- Display Waveform
- Maintain Arterial
Line Alarms
|
- Maintain Closed System
- Obtain Blood Samples
- Change Dressings
- Monitor Arterial
Site
- Document
- Remove Arterial
Line
|
| STANDARD
OF NURSING CARE |
RATIONALE
FOR STANDARD |
| 1. |
Set-Up
Hemodynamic Circuit
Clinical Nurses in CCTC are responsible
for the priming, zeroing, leveling, and maintenance of hemodynamic
pressure monitoring circuits and for the assessment and monitoring
of hemodynamic pressures and waveforms.
Clinical Nurses may flush hemodynamic monitoring
circuits as required to maintain patency.
|
1. |
Clinical Nurses in CCTC have received additional education in the
care of invasive monitoring circuits. Refer to Procedure for Hemodynamic
Monitoring Circuits.
|
| 2. |
Maintain Accuracy
Hemodynamic transducers are zeroed at each initial setup, with
the air-fluid interface (stopcock above transducer) leveled to the
mid-axillary line.
Transducer levels should be validated at
the beginning of each shift, prior to each PWP or CVP measurement,
following position changes, and prn to validate hemodynamic pressures.
Document level confirmation in the graphic
record. Documentation is required at the start of each shift and
q6h, prior to each CVP and PWP measurement, and following transducer
repositioning.
|
2. |
Zeroing eliminates the effect of atmospheric pressure on measured
hemodyamic values.
Leveling eliminates the influence
of hydrostatic pressure on the transducer. A transducer that is
positioned below the patient's heart will produce falsely elevated
pressures and a transducer positioned above the patient's heart
will produce falsely low pressures. |
| 3. |
Monitor Blood Pressure
Continuous arterial pressure monitoring is indicated for patients
requiring BP monitoring >q1h, receiving continuous IV infusion
of medications that affect cardiac output/blood pressure, requiring
frequent blood gas monitoring or who are hemodynamically/neurologically
unstable.
Prior to recording hemodynamic values, evaluate
pressure waveforms, verify transducer levels, and ensure that sufficient
flush volume and pressure is present.
When comparing NIBP to arterial line pressures, MBP provides the
most appropriate comparison. Compare cuff to NIBP pressures taken
from the same limb. |
3. |
Cuff BP measurements become increasing less accurate when hypotension
develops. Accuracy is also affected by user technique and cuff size.
Insertion of an arterial line is easiest when a patient has an adequate
BP. Continuous infusion of vasoactive drugs necessitates continuous
BP monitoring to evaluate response to drug therapy and to identify
adverse effects.
NIBP cuffs do not provide continuous
pressure monitoring. Although more accurate than auscultative BP
monitoring, NIBP cuffs can become inaccurate when peripheral circulation
is compromised. NIBP cuffs are uncomfortable and can produce bruising
in susceptible patients; they are not appropriate when >q1h BP
monitoring is required. Repetitive inflation of a BP cuff can lead
to falsely elevated readings.
Systolic BP readings obtained
from invasive arterial catheters are often accentuated, with reduced
diastolic readings. NIBP measurements utilize different technology
and may produce systolic readings that do not match the arterial
line systolic reading. MBP from arterial lines and NIBP circuits
provide more appropriate correlation.
Patients may have differing
BPs between left and right limbs, and peripheral circulation may
be reduced in shock.
Arterial pressures may be falsely
elevated if an artery is in spasm. This can occur following repeated
insertion attempts. A warm towel wrapped around the site may help
to reduce spasm.
|
| 4. |
Display Waveforms
Invasive arterial lines and right heart
catheters must be connected to a monitor that provides a continuous
waveform display. |
4. |
Waveform quality must be assessed prior to acceptance of a displayed
pressure value. Pressures are derived from the detected waveform;
poor quality waveforms will produce inaccurate pressure readings.
The waveform from the tip of
the right heart catheter must be displayed continually to demonstrate
the catheter tip location within the heart. Continuous monitoring
is required to promptly detect catheter migration into the right
ventricle or pulmonary wedge position.
|
| 5. |
Maintain
Arterial Line Alarms
Appropriate alarms
must be on for all patients requiring continuous arterial pressure
monitoring. Alarm settings should be selected based on the degree
of fluctuation in the patient's BP. Upper and lower alarm limits
that represent clinically important changes are selected for each
individual patient.
High and low alarm settings must be assessed and documented
each hour in the graphic record.
Alarms may need to be disabled if an arterial
line becomes positional. If an arterial line alarms is disabled,
documentation in the AI record is required. Documentation should
include the reason for disabling the alarm and should describe troubleshooting
strategies.
|
5. |
Alarms
provide prompt notification of both changes in the BP and accidental
disconnection/opening of the circuit. |
| 6. |
Maintain
Closed System
All stopcocks must have dead-end (non-vented) luer lock caps or
luer lock connected infusions. This includes stopcocks located on
transducers.
Hemodynamic circuits are changed with
each new line and prn.
|
6. |
To prevent accidental entry of air or contaminants. Accidental movement
of a stopcock that does not have a dead end luer lock cap can result
in hemorrhage/air embolus.
|
| 7. |
Obtain Blood Samples
Clinical Nurses may draw blood from indwelling arterial and venous
lines. All infusion must be stopped (introducer, distal and proximal
infusions from multiple lines) during blood withdrawal. Stopcocks
should be turned to 45 degrees between syringe changes. Clearlink (TM) access caps should be changed when drawing blood culture samples and a minimum of q 7 days and prn.
|
7. |
See Procedure for
Blood Withdrawal; Blood
Gases Arterial; and Blood
Gases Venous. |
| 8. |
Change Dressings
Arterial line dressings are changed q 2 days (tape and gauze)and prn, when using Tegaderm(TM) transparent dressing change q 7days and prn when occlusivity
is disrupted. Document date of change in the graphic record and
kardex. Record any redness or abnormal findings in the AI record
and report to the physician. See Procedure
for Arterial and Central Line Dressings. Don appropriate PPE when accessing an arterial line.
Lines are changed prn for evidence of site
infection or sepsis.
|
8. |
Dressings of tape and gauze are change q 2 days and prn in critically ill patients to allow
site inspection and to reduce bacterial colonization Skin condition, diaphoresis and site oozing necessitates more frequent
dressing changes in critically ill patients.
It is acceptable and prefered LHSC Infection Control practice to use a Tegaderm (TM) dressing that
can be changed q7days as per general hospital policy, as long as
occlusivity is maintained.
The Centre for Disease Control
does not recommend routine line changes.
|
| 9. |
Monitor Arterial Site
Arterial line sites/dressing should be kept as visible as possible.
Check the site q1h and prn to assess for bleeding.
Use minimal dressing material.
Assess distal extremity for evidence of compromised
color, circulation or motion q1h. |
9. |
Bleeding can occur very rapidly from dislodged arterial catheters
or opened circuits.
Bulky dressings can mask bleeding.
Pressure dressings will not stop arterial bleeding.
Thrombus, catheter spasm and
small vessels can lead to compromised distal circulation.
|
| 10. |
Document
Record S/D and M BP in the graphic record q1h and prn.
Assess site q shift and document in the 24
hour assessment record. Assess patency q shift and q 4 h and document
in the intravascular device section of the flow sheet.
Sign for heparinized flush solution in the
MAR.
Document dressing changes in flow sheet.
Star and DAR abnormal findings.
|
10. |
Per hospital standards.
|
| 11. |
Remove Arterial Line
Clinical Nurses
may remove arterial lines in CCTC, with an order from a physician.
Document removal in the intravascular device section of the graphic
record.
See Procedure
for Removal of Peripheral Arterial Line or Femoral
Arterial Line.
|
11. |
|